Glossary
A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
D
DDDR. NBG code for dual-chamber pacing with atrial tracking
and rate modulation.
DDDRD. NBG code for DDDR pacing with the added capability
of dual antitachycardia responses, both antitachycardia pacing and
delivery of a high-energy shock (cardioversion and defibrillation).
DDI. NBG code for dual-chamber, nontracking pacing, with
sensing in both chambers. Pacing and sensing occur in both atrium
and ventricle, with inhibited response.
DDIR. The NBG code for dual-chamber, nontracking pacing
with sensing in both chambers and rate modulation.
Defibrillation. The termination of atrial or ventricular
fibrillation, usually using electrical energy delivered to the heart
through chest-wall electrodes, directly to the heart during open-heart
surgery, or through an automatic implantable defibrillator.
Demand. Any pacing mode or pacemaker that delivers an output
pulse only when necessary, that is, when the intrinsic rate is less
than the programmed base rate.
Demand interval. The time period between two consecutive
paced events in the same chamber without an intervening sensed event.
It is expressed in milliseconds (AA interval, VV interval). Also
known as basic interval or pacing interval. demand rate. The rate
(in ppm) defined by the demand interval (in milliseconds).
Depletion indicator. A function of the pacemaker showing
that the battery is wearing out and pulse generator replacement
should be scheduled.
Depolarization. The sudden change in electrical potential
from negative to slightly positive which occurs during phase O of
the action potential. In the heart, electrical depolarization initiates
the mechanical contraction. Waves of depolarization spread from
cell to cell. When this occurs in the atria, a native P wave is
seen on the EGG; when this occurs in the ventricles, this is seen
as a native QRS complex on the EGG. Depolarization can also be initiated
by a pacemaker output stimulus.
Diaphragmatic stimulation. Electrical activation of the
diaphragm muscle by the pacemaker output pulse. The abrupt diaphragmatic
contraction is noted clinically as hiccups associated with each
pacing stimulus. The pacing stimulus may stimulate the diaphragm
either directly or indirectly via the phrenic nerve.
Distal. Farthest from the point of reference. On pacing
lead connectors, the distal electrode is the tip electrode. See
also proximal.
DOO. The NBG code for dual-chamber asynchronous pacing.
DOOR. The NBG code for dual chamber asynchronous pacing
with rate modulation.
Dual chamber. Refers to a pacing mode or pulse generator
capable of pacing and/or sensing in the atrium and pacing and/or
sensing in the ventricle. Examples of dual-chamber modes are VAT,
VDD, DVI, DDI and DDD.
Dual Chamber Pacing. Pacing in both the atria and
the ventricles to artificially restore the natural contraction sequence
of the heart. Also called ‘physiologic” or “universal” pacing.
DVI. The NBG code for AV sequential pacing. Pacing occurs
in the atrium and ventricle, but sensing occurs only in the ventricle
and the response to sensing is inhibition.
DVIR. The NBG code for AV sequential pacing with rate modulation.
Dysrhythmia. Any disturbance in the normal cardiac rhythm,
possibly resulting from abnormal origination, discharge or conduction
of the electrical impulses. In common usage, it is synonymous with
arrhythmia, although in the strictest sense, arrhythmia is the absence
of cardiac rhythm and dysrhythmia is an abnormal cardiac rhythm. |