Glossary
A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
F
Fallback. A programmable option in some pacemakers
designed to handle rapid atrial rates.
Far-field signal. On an intracardiac electrogram, the intrinsic
signal originating from the opposite chamber of the heart from the
one in which the lead electrode is located. For example, in an atrial
electrogram, the QRS wave is the far-field signal, but in a ventricular
electrogram, the P wave is the far-field signal.
Feedthrough. The insulated electrical pole(s) in the pulse
generator connector that serves to connect the pulse generator's
internal circuitry to the pacing lead. Feedthroughs are hermetically
sealed to prevent leakage of fluids into the pulse generator and
are typically made of glass or ceramic material.
Fibrillation. A type of cardiac arrhythmia characterized
by rapid, unsynchronized quivering of atria or ventricles. Atrial
fibrillation may be asymptomatic, but ventricular fibrillation is
typically fatal if not corrected within minutes.
Fibrillatory wave. The erratic baseline waves on an ECG
during fibrillation. Abbreviated f wave.
Fibrosis. The buildup of scar tissue.
Fibrotic capsule. Scar tissue which forms around the pacing
electrode tip or the pulse generator. This is the body's normal
reaction to any foreign material. first-degree AV block. A type
of AV block characterized by an increase in the time required for
a P wave to be conducted to the ventricle. The PR interval is greater
than 200 ms. Sometimes called first degree heart block.
Fixation. The means by which a pacing lead is attached to
the heart, usually subdivided into active fixation (using a mechanism
such as a screw-in tip) and passive fixation (using tines or fins).
Fixed rate pacing. A type of pacing in which output pulses
are delivered at a preset rate with no modification from sensing
or other input. Also known as asynchronous pacing.
Fluoroscope. The device used routinely during lead placement
which provides real-time X-ray images on a monitor, allowing the
physician to observe the transvenous passage and placement of the
lead into the heart. Fluoroscopes may be further defined by the
angle at which they take images: AP fluoroscopy (anterior-posterior)
and/or lateral fluoroscopy are used for implantation procedures.
Flutter. A rapid but regular rhythm (250 to 350 bpm in the
atria or 200 to 300 bpm in the ventricles) often seen on the ECG
as a saw-toothed pattern. Ventricular flutter may result in death
unless corrected in minutes.
Flutter wave. The rapid, regular, uniform saw-toothed waves
on an ECG during atrial flutter. Abbreviated F wave.
Frequency. The number of times a signal occurs in a specific
period of time. Filters in the pulse generator's sense amplifiers
selectively screen out signals at frequency ranges not associated
with signals of cardiac origin.
Fusion beat. A cardiac depolarization (atrial or ventricular)
resulting from two foci. In pacing, it typically refers
to the ECG waveform which results when an intrinsic depolarization
and a pacemaker output pulse occur simultaneously and both contribute
to the electrical activation of that chamber. See also pseudofusion
beat. |