Glossary
A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
P
P wave. (1)An intrinsic atrial event. (2) The portion of
the ECG representing atrial depolarization.
Pacemaker. (1) Devices. The system composed of the pulse
generator and the connected lead(s). In common usage, pulse generator
is synonymous with pacemaker. (2) Physiology. Any portion of the
heart which initiates the electrical impulses that control the heart
rate; in the healthy heart, this is the SA node. In pacing, a physiological
pacemaker is often called a natural pacemaker.
Pacemaker-dependent patient. A patient who needs a pacemaker
to maintain proper cardiac rhythm to sustain adequate cardiac output
sufficient for the activities of daily living. Such an individual
may have little or no underlying intrinsic rhythm.
Pacemaker EGG. An ECG from a patient being artificially
paced. Pacemaker ECGs are usually characterized by the presence
of pacemaker spikes.
Pacemaker-induced arrhythmia. A type of dysrhythmia, usually
a tachycardia, induced by competition between the paced and native
rhythms. Its induction may be inadvertent or intentional, as in
an EP lab. Once initiated, the pacemaker-induced arrhythmia is sustained
without the continued participation of the pacemaker.
Pacemaker-mediated tachycardia. A tachycardia induced by
competition between the paced and native rhythms and sustained by
the continued participation of the pacemaker. A common example of
pacemaker-mediated tachycardia is an endless loop tachycardia. and,
rarely, intrinsic loops within the pacemaker. Abbreviated PMT. Also
known as endless loop tachycardia, pacemaker reentry tachycardia
or circus tachycardia.
Pacemaker Mediated Tachycardia (PMT). A reentrant arrhythmia
in which the dual chamber pacemaker acts as the anterograde limb
of the tachycardia and the natural conduction path as the retrograde
limb. A rapid paced rhythm that can occur with atrial tracking pacemakers.
It begins with the loss of A-V synchrony and is sustained by ventricular
events that are conducted retrogradely to the atria. The pacemaker
senses this retrograde atrial depolarization and then delivers a
stimulus to the ventricle. Also known as endless less loop tachycardia.
Other examples include tracking atrial fibrillation or flutter
Pacemaker reentry tachycardia. circus tachycardia or pacemaker
mediated tachycardia.
Pacemaker syndrome. Adverse hemodynamic or electrophysiologic
consequences, or both, associated with an electrically normal pacing
system. Pacemaker syndrome may induce symptoms ranging from minor
to severe or may simply limit the patient's ability to achieve an
optimal functional status. While most commonly associated with the
loss of AV synchrony, it may also occur because of an inappropriate
AV interval or inappropriate rate modulation. It may be intermittent
or persistent, but will only be present during a paced rhythm.
Pacing configuration. The polarity (unipolar or bipolar)
with which a pacemaker will stimulate. Many modern pulse generators
offer programmable pacing configuration. Sometimes called pulse
configuration. See also programmable polarity.
Pacing interval. The time period between two consecutive
paced events without an intervening sensed event. It is expressed
in milliseconds (AA interval, VV interval) Also known as automatic
interval, demand interval or basic interval.
Pacing state. Refers to the primary states (or event combinations)
of DDD pacing. These are PV (sensed P wave followed by a ventricular
output pulse), PR (sensed P wave followed by a sensed R wave), AV
(atrial pulse followed by a ventricular output pulse), AR (atrial
output pulse followed by a sensed R wave). PVE (premature ventricular
event) is sometimes included as a pacing state.
Pacing stimulus. (1)The output pulse from the pacemaker.
(2) The mark representing the pacemaker's electrical output recorded
by the EGG. Also known as output pulse. See also artifact.
Pacing system. The implanted pulse generator, implanted
lead(s) and programmer, taken together as one system, plus the patient.
Pacing system analyzer. An external testing and measuring
device capable of pacing the heart during pacemaker implantation
and measuring such things as the stimulation and sensing thresholds
and lead impedance. Pacing system analyzers may also have the capability
of testing pulse generator function prior to implant, measuring
slew rate or printing an electrogram of a sensed QRS complex. Abbreviated
PSA.
Parameter. A distinguishable value that can be measured
and sometimes changed, either indirectly or directly. In pacing,
parameter refers to a value that influences the function of the
pacemaker, such as mode, base rate, pulse width, pulse amplitude,
sensitivity and so on.
Parasympathetic nervous system. Part of the autonomic nervous
system involved in the control of involuntary bodily functions,
including control of cardiac and blood vessel activity. Stimulation
of certain parasympathetic nerves causes the SA node to discharge
at a slower rate and slows conduction through the AV node, which,
in turn, extends the PR interval. See also autonomic nervous system
and sympathetic nervous system.
Paroxysmal. Pertaining to a sudden, periodic recurrence
or worsening of disease symptoms.
Paroxysmal supraventricular tachycardia. Type of rapid atrial
arrhythmia characterized by brief periods of sudden-onset and often
abrupt termination of atrial tachycardia. The sudden onset of the
tachycardia is caused by micro-reentry within the AV node or macro-reentry
between the AV node and a bypass tract. Abbreviated PSVT. Sometimes
called paroxysmal atrial tachycardia, AV nodal reentry tachycardia
or AV reentry tachycardia.
Passive fixation lead. A pacing lead with an attachment
mechanism such as fins or tines which allow them to lodge in the
trabeculae lining the inner wall of the heart without actually penetrating
the cardiac wall. See also active fixation lead.
Peak amplitude. A general term for the greatest amplitude
of a waveform, measured from baseline to highest point of wave.
percutaneous. Literally "through the skin." In pacing,
endocardial leads are inserted through the skin into a blood vessel,
then advanced into the heart.
Percutaneous introducer. A thin, hollow tube used for easy
passage of the endocardial lead through the skin into a blood vessel,
then advancing the lead into the heart.
Permanent lead. Any implanted pacing lead intended for long-term
use.
Phantom programming. A general term for any inadvertent
or unintentional programming of a pulse generator. Phantom programming
may be caused by electromagnetic interference in the environment
affecting programming or it may be due to an intentional reprogramming
of the pulse generator which had not been recorded in the patient's
record.
Phrenic nerve. The nerve controlling the rhythmic contraction
of the diaphragm.
Physiological refractoriness. The normal period during which
cardiac tissue will not respond to a stimulus, meaning that cardiac
tissue will not conduct or contract during this time.
Pocket. The subcutaneous pouch in which the pulse generator
is inserted at implant.
Pocket stimulation. Stimulation of muscle tissue surrounding
the implanted pulse generator. Pocket stimulation may occur in a
unipolar system owing to the concentration of current flow at the
pulse generator case (the anode). Another potential cause of pocket
stimulation is a defect in the lead insulation, which could occur
with either a unipolar or a bipolar pulse generator.
Polarity. Pertaining to the electrode configuration of a
lead or the operating configuration of a pulse generator. Polarity
is either unipolar or bipolar. It is possible in some advanced programmable
polarity pulse generators to pace in one polarity and sense in another.
Note that a bipolar lead (having two electrodes) can pace or sense
in either a unipolar or bipolar manner, while a unipolar lead (having
one electrode only) can pace or sense only in a unipolar manner.
Polymorphic tachycardia. A tachycardia that originates from
more than one focus. On an EGG, the QRS complex will exhibit a variety
of different morphologies.
Polyurethane. A synthetic resin polymer used as insulation
material in some pacing leads.
Post-ventricular atrial refractory period (PVARP). One of
the timing cycles in dual-chamber pacemakers during which the atrial
channel's sensing circuitry is rendered unresponsive. Initiated
by a sensed or ventricular paced event and ending with an atrial
sensed or paced event, the post-ventricular atrial refractory period
is intended to prevent the atrial sense amplifier from sensing farfield
ventricular signals. Therefore, the post-ventricular atrial refractory
period may be considered the atrial equivalent of the ventricular
blanking period. The post-ventricular atrial refractory period is
usually a programmable parameter; if sufficiently long, it can prevent
the atrial channel from responding to premature atrial contractions
or retrograde atrial depolarizations and thus prevent pacemaker-mediated
tachycardias. Abbreviated PVARP.
PPM. Abbreviation for pulses per minute. The abbreviation
ppm usually refers to the paced rate, while beats per minute (BPM)
refers to an intrinsic heart rate. See also bpm.
PR. One of the four normal pacing states in DDD and DDDR
pacing, starting with an atrial sensed event (P) and ending with
a ventricular sensed event (R). See also pacing state.
PR interval. (1)The interval corresponding to the PR pacing
state. (2) The period of time, measured in milliseconds and readily
viewed on a surface EGG, from the onset of the P wave (atrial depolarization)
to the onset of the QRS complex (ventricular depolarization).
Presyncope. The sensation of almost fainting but without
total loss of consciousness. Sometimes called near-syncope.
Programmability. The characteristic of most modern pulse
generators allowing one or more parameters to be adjusted noninvasively
before, during or after implantation. Also known as multiprogrammability.
Programmable polarity. A feature of pacemakers in which
the polarity configuration for pacing and/or sensing (unipolar or
bipolar) may be programmed. In some pacemakers, it is possible to
program sensing and pacing polarity independently (i.e. to pace
unipolar but sense bipolar or vice versa).
Proximal. Nearest the point of reference. On pacing lead
connectors, the proximal electrode is the ring electrode. See also
distal. pseudofusion beat. A spontaneous cardiac depolarization
occurring at or near a pulse generator output pulse. Because the
stimulus occurs after the heart has spontaneously depolarized, the
pacemaker output is ineffective, but it distorts the morphology
of the complex on the EGG. See also pseudopseudofusion beat.
Pseudomalfunction. (1)An unexpected behavior of the pacemaker
which may appear to be the result of a malfunction but which is
actually normal behavior for that device under those programmed
settings. Pseudomalfunctions are often eccentric or unusual behaviors.
(2) Normal pacing system function which is misinterpreted as a pacing
system problem due to a recording system problem or artifact.
Pseudopseudofusion beat. An electrocardiographic superimposition
of an atrial stimulus on the native QRS complex. The atrial output
pulse cannot contribute to the ventricular contraction. Pseudopseudofusion
beats can occur only in dual-chamber pacing modes and are most common
in DVI. See also pseudofusion beat.
Pulse amplitude. The height of the output pulse as expressed
in volts or milliamperes.
Pulse duration. See pulse width.
Pulse generator. The implantable device that contains the
battery, circuitry and other components which provide the output
stimuli. The pulse generator and the lead(s) together make up the
pacemaker, although in informal use, pulse generator and pacemaker
are used as synonyms.
Pulse pressure. The difference between the systolic and
diastolic blood pressures.
Pulse width. The length of time, measured in milliseconds,
that the output pulse is delivered. Pacemaker output is usually
defined in terms of voltage (pulse amplitude) and milliseconds (pulse
width). Pulse width is measured from the leading edge to the trailing
edge of the wave. Also known as pulse duration.
Pulses per minute. The standard unit of measure for paced
rate. Abbreviated ppm.
PV. One of the four normal pacing states in DDD and DDDR
pacing, starting with an atrial sensed event and ending with a ventricular
paced event. See also pacing state.
PV interval. The time period from the onset of the P wave
(atrial depolarization) to the ventricular pacing stimulus.
PVARP. Abbreviation for post-ventricular atrial refractory
period. |