Glossary
A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
S
Safety pacing. A mechanism available in certain dual-chamber
pacemakers which forces a ventricular output when an event is sensed
during a special interval (called the ventricular safety standby
in Pacesetter devices) following the ventricular blanking period.
The purpose of safety pacing is to eliminate crosstalk inhibition.
See also ventricular safety standby.
Second-degree heart block. A condition characterized by
the failure to conduct some, but not all, electrical impulses from
the atria to the ventricles. This blockage can take different forms:
Mobitz I or Wenckebach (PR interval lengthens progressively until
a P wave is not conducted) and Mobitz II (constant PR interval with
blocked beats).
Sense amplifier. The electronic component in the pacemaker's
sensing circuit which detects and amplifies incoming signals for
processing.
Sensing. The ability of the pacemaker to recognize and respond
to electrical activity in the heart. How the pacemaker responds
to sensed signals depends on its programmed mode and parameters.
Sensing configuration. The polarity (unipolar or bipolar)
with which a pacemaker will sense. Many modern pulse generators
offer a programmable sensing configuration. Some Pacesetter pacemakers
offer three polarity options for sensing: unipolar tip (from the
tip electrode to the pulse generator case), unipolar ring (from
the ring or proximal electrode to the pulse generator case) and
bipolar (from the tip electrode to the ring electrode). See also
programmable polarity.
Sensing threshold. The largest intrinsic atrial or ventricular
signal, expressed in mV, which can be consistently sensed by the
pulse generator's sense amplifier. The sensing threshold is assessed
by measuring the R wave or P wave, respectively, for ventricular
or atrial sensing.
Sensitivity. A pacemaker parameter which determines the
amplitude of signals to which the device's sense amplifiers will
respond. Sensitivity is stated in millivolts. Note that the higher
the millivolt value, the lower the sensitivity. (If sensitivity
is 6 mV, a signal has to be 6 mV or larger before the signal is
recognized; on the other hand, if sensitivity is 2 mV, a signal
only has to be 2 mV or larger before it is recognized. Thus the
lower the mV value, the more sensitive the device.)
Sensor. In rate-modulated pacing, a device or component
that attempts to assess the patient's need for rate support by monitoring
a condition other than intrinsic atrial activity and translating
this information into a corresponding increase or decrease of programmed
base rate. Typical conditions evaluated by sensors are vibrations
(relating to activity), central venous blood temperature, minute
ventilation or oxygen.
Sensor threshold. The programmable value for rate-modulated
pacemakers from which the sensor level is derived. In general, the
higher the sensor threshold, the higher the level of activity needed
to bring about an initial increase in pacing rate above the base
rate.
Set-screw. A type of screw in the pulse generator's epoxy
connector used to tighten the lead's terminal pin in the receptacles
and press it securely against the connector block.
Sick sinus syndrome. See sinus node dysfunction. Abbreviated
SSS.
Silicone. An organic plastic compound used as insulation
on some pacing leads.
Single-chamber pulse generator. A pulse generator that paces
and senses only one cardiac chamber. Single-chamber pulse generators
may be atrial or ventricular; it is often possible to use the same
pulse generator for either atrial or ventricular application.
Single-pass lead. A pacing lead designed to be passed transvenously
and lodge in the ventricle but in such a way that it places one
or more electrodes in both the atrium and the ventricle. A single-pass
lead allows pacing and sensing in the ventricle, but can sense only
in the atrium.
Sinoatrial block. The failure of electrical impulses produced
by a normally functioning SA node from reaching the atria. There
are three degrees of sinoatrial block. Abbreviated SA block.
Sinoatrial node. A group of cells located high and posterior
in the right atrium at the point where the superior vena cava joins
the atrial tissue mass which acts as the natural pacemaker in the
healthy heart. The healthy sinoatrial node initiates 60 to 100 beats
per minute at rest. Abbreviated SA node.
Sinus node dysfunction. A collective term for any of several
sinus node dysrhythmias including marked sinus bradycardia, SA block,
SA arrest and bradycardia-tachycardia syndrome. AV nodal function
is usually normal, thus the P wave is usually conducted. Sometimes
called sick sinus syndrome or sinoatrial node dysfunction..
Slew rate. In a segment of the intracardiac electrogram,
the amount of change in signal amplitude (voltage) divided by the
period of time in which the change occurred. It appears graphically
as the steepest slope of the waveform. Slew rate is usually expressed
in mV per ms or volts per second. It is mathematically defined as
dV/dt. Most pulse generator sense amplifiers require a slew rate
of greater than or equal to 0.5 V/sec in order for a signal to be
sensed.
Slope. The programmable parameter in rate-modulated systems
which determines the increase in pacing rate above the base rate
that will occur at different levels of sensor input signals. See
also Auto-Set.
Spike. A small but sharply vertical deflection that appears
on the surface ECG indicating that a pacemaker output was delivered.
It is caused by the brief discharge of electricity produced by the
pacemaker to stimulate the heart. In some situations, a pacemaker
spike may not appear clearly on an EGG. See also artifact.
SSI. A manufacturer's code for single chamber inhibited
pacing. Pacing and sensing occur in either atrium or ventricle with
inhibited response. Once connected to a pacing lead, the mode is
stated AAI (if the lead is in the atrium) or VVI (if the lead is
in the ventricle).
SSIR. SSI pacing with sensor ON, providing rate-modulated
pacing during physical activity.
ST segment. The portion of the ECG representing the end
of ventricular depolarization (end of the R wave) and the beginning
of ventricular repolarization (T wave).
Stimulation threshold. The minimum output from the pacemaker
which consistently elicits a cardiac depolarization and contraction.
Stimulation thresholds may be measured as voltage, current, pulse
width, energy or charge. Sometimes called capture threshold or pacing
threshold.
Stimulus. See output pulse.
Stokes-Adams syndrome. Paroxysmal heart block or ventricular
dysrhythmia, producing low or no cardiac output resulting in such
symptoms as syncope or possibly convulsions.
Strength-duration curve. The curve reflecting how much threshold
voltage (pulse amplitude) and duration (pulse width) are necessary
to capture the heart. This curve demonstrates the interaction of
pulse amplitude and pulse width. At a point on the graph, usually
about 1 ms, the curve flattens out to show the rheobase or the point
where progressive increases in pulse width no longer affects the
voltage threshold. See also rheobase, chronaxie.
Stroke volume. The volume of blood ejected per heart beat.
Stroke volume equals end-diastolic volume minus end-systolic value.
Supraventricular tachycardia. A tachycardia that originates
from the supraventricular region. Supraventricular tachycardias
are usually pathologic tachycardias and include AV and AV nodal
reentry tachycardia. Abbreviated SVT.
Surface electrocardiogram. A graphic depiction of the electrical
signal emitted by active cardiac tissue recorded through electrodes
from the body surface.
Sutureless lead. A type of epicardial pacing lead which
attaches to the myocardium with an active fixation mechanism rather
than a suture. An example of a sutureless lead is the "corkscrew"
type active fixation pacing lead.
Sympathetic nervous system. Part of the autonomic nervous
system involved in the control of involuntary bodily functions,
including the control of cardiac and blood vessel activity. This
system stimulates cardiac activity and produces effects opposite
those of the parasympathetic nervous system, which depresses cardiac
activity. Some effects of sympathetic stimulation are an increase
in heart rate, cardiac output and blood pressure. See also autonomic
nervous system and parasympathetic nervous system.
Synchronous. A type of pacing in which output pulses are
delivered in response to a sensed event, usually after an appropriate
delaying timing cycle. For example, atrial synchronous pacing provides
atrial pacing in response to sensed atrial activity. Examples of
synchronous modes are AAT, AATR, VVT, WTR, DDD and DDDR.
Syncope. Fainting or a black-out spell caused by insufficient
blood supply to the brain.
|