Glossary
A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
U
Underdrive pacing. Pacing at a rate slower than a tachycardia,
which can sometimes interrupt and terminate a reentrant arrhythmia.
Underlying rhythm. The basic intrinsic rhythm on which the
paced rate or certain arrhythmias may be superimposed, e.g. sinus
arrest, sinoatrial (SA) exit block, atrial fibrillation or AV block.
Undersensing. The failure of the pacemaker to sense P or
R waves with the result that inappropriately timed, asynchronous
or competitive output pulses are delivered. Undersensing can sometimes
be corrected by programming the pacemaker to a more sensitive setting,
i.e., decreasing the mV value. See also oversensing.
Unipolar. Having one pole. In pacing, unipolar may refer
to a lead with a single electrode or a pulse generator that can
accommodate such a lead. Unipolar is actually a misnomer, in that
the circuit has two poles (the pulse generator case serves as the
other pole) even if the lead itself does not. See also bipolar.
Unipolar lead. A pacing lead with a single electrical pole
at the distal tip of the pacing lead (negative pole). The anode
(positive pole) is the pulse generator case. The cathode is the
electrode through which the stimulating pulse is delivered. See
also bipolar lead.
Unipolar pulse generator. A pulse generator that can accommodate
a unipolar lead and operate in a unipolar pacing and sensing configuration.
Upper rate. In DDD pacemakers, this is the programmed limit
to which the ventricles may be paced in response to sensed atrial
activity. Also see maximum tracking rate. |